Abstract

A temporally- and spatially-resolved X-ray spectrometer,with spectral range of 1 500~2 300 eV and temporal resolution about 2 ns,has been developed for diagnosing K-shell radiation of aluminum wire array Z-pinch plasma driven by Qiangguang-I generator (1.5 MA,80~100 ns).It consisted of an elliptically-curved TAP crystal and a four-channel X-ray frame camera.Theoretical analyses were made to check out possible effects of different signal transfer and conversion processes on spectral response of the spectrometer.It recorded time-resolved and -integrated K-shell resonance lines of H-,He-like aluminum ions and satellite lines of Li-,He-like ions.Results were compared with spectral-resolved X-ray images obtained by pinhole cameras.Time-resolved K-shell spectra show many differences on line ratios and axial distribution with each other and time-integrated results,which exhibits variations of electron temperature and ion population on a 5~10 ns time scale.

关键词

Z箍缩/K层辐射/弯晶谱仪/分幅相机

Key words

Z-pinch/K-shell radiation/curved crystal spectrometer/frame camera

分类

数理科学

引用本文复制引用

吴刚,雷天时,韩娟娟,孙铁平,邱爱慈,吕敏,黑东炜,盛亮,魏福利,蒯斌,王亮平,丛培天..用于测量Z箍缩铝等离子体K层辐射的邻苯二甲酸氢铊弯晶摄谱仪[J].强激光与粒子束,2009,21(7):1115-1120,6.

基金项目

国家自然科学基金重点项目(10635050) (10635050)

国家自然科学基金项目(10475065) (10475065)

强激光与粒子束

OA北大核心CSCDCSTPCD

1001-4322

访问量2
|
下载量0
段落导航相关论文