检验医学与临床2009,Vol.6Issue(14):1159-1160,1162,3.
同室不同生化检测系统测定结果的比对和偏倚评估
The method of the comparison and bias estimation on the result in different biochemical testing system under the same laboratory condition
梁瑞莲 1周远青 1谢丽明 1唐跃华 1卓伟 1罗建伟1
作者信息
- 1. 南方医科大学附属顺德第一人民医院检验科,广东顺德,528300
- 折叠
摘要
Abstract
Objective The comparability of serum results of different biochemical detecting system were studied through the method of comparison and bias estimation between Hitachi 7170 and Olympus AU1000.Methods According to the profile NCCLS EP9-A2, two different biochemical testing system,Hitachi 7170(comparison method) and Olympus AU1000(laboratory method),were used respectively to test 16 serological items to obtain the correlation coefficient and the linear equation, and the system bias % at the level of medical decision were calculated.The comparability of different biochemical testing system was judged according to half of the clinical laboratory improvement amendment 88(CLIA`88) 1/2 of the total errors is allowed.Results The testing results of the different biochemical testing system had a great correlation, All correlation coefficients of 16 items were greater than 0.975. The system bias of ALB and UA at a level of medical decision were not acceptable , and the rest could be accepted.Conclusion If the same item is tested in more than two different systems,it is necessary to judge clinical acceptability by the method of comparison and bias evaluation so as to insure the comparability.关键词
全自动生化分析仪/偏差评估/方法比对/医学决定水平Key words
automatic biochemical analyzer/bias estimation/method comparison/the level of medical decision分类
临床医学引用本文复制引用
梁瑞莲,周远青,谢丽明,唐跃华,卓伟,罗建伟..同室不同生化检测系统测定结果的比对和偏倚评估[J].检验医学与临床,2009,6(14):1159-1160,1162,3.