Abstract
Objective To establish a simple and sensitive a utomatic methodof determination of copper in serum with two reagent. Method For determination of serum copper with 5-Br-PADAP in the presence of mixed surfactant.Results The method was liner in the linar range of the method was 0~80μmol/L. The average rate of recovery was 100.5%.The within-run and between-run CV were 0.019and 0.024.Comparing this method(Y) with AAS(X),we obtained the regression equation.(Y=0.9910X+0.0923,r=0.9923 and P>0.05).The reference range(±2s) was 11.52~21.96μmol/L(n=120).Conclusion This method is simple,sensitive and suitable for clinical application .关键词
铜/血清/自动分析法/5-Br-PADAP分类
医药卫生引用本文复制引用
连国军,李东,陈红,方芳..5-Br-PADAP自动化分析法测定血清铜[J].陕西医学检验,2001,16(1):12-13,2.