首页|期刊导航|哈尔滨工业大学学报(英文版)|The experimental study of the calibration of the focus ultrasonic transducer on its picture
哈尔滨工业大学学报(英文版)2005,Vol.12Issue(2):139-142,4.
The experimental study of the calibration of the focus ultrasonic transducer on its picture
The experimental study of the calibration of the focus ultrasonic transducer on its picture
摘要
Abstract
The shape and size of the energy transducer' s focus decide the constitution of the cure project using HIFU ( High intensity focused ultrasound). The parameter of the focus will be different when the structure of the transducer is different, so we must calibrate each of them. This paper represents calibration principle and method of focal territory in a high-power ultrasonic focus transducer. The experiment is performed on our designed multiple wafer concave focus probe and HIFU apparatus, using 1010 nylon and pork as the target range.It can be seen that the geometric center and focal territory acoustic field center cannot match together. When using nylon as the target, the distributed curves and theoretical curves are a better match. Therefore, the method and principle of calibration here is constructive for the calibration on pictures of other types of HIFU apparatus.关键词
ultrasonic wave/the focus ultrasonic transducer/focus/calibration/HIFUKey words
ultrasonic wave/the focus ultrasonic transducer/focus/calibration/HIFU分类
医药卫生引用本文复制引用
HOU Zhen-xiu,ZHAO Yong-ping,HUANG Nai-hua,CHEN Xin-liang,ZHANG Li-li,LUO Jian-wei,JIN Chang-shan..The experimental study of the calibration of the focus ultrasonic transducer on its picture[J].哈尔滨工业大学学报(英文版),2005,12(2):139-142,4.基金项目
Sponsored by the Natural Science Foundation of Heilongjiang Province( Grant No. F0205). ( Grant No. F0205)