| 注册
首页|期刊导航|中华医学杂志(英文版)|Clinical study of domestic ZQL-type retrievable vena cava filter

Clinical study of domestic ZQL-type retrievable vena cava filter

XU Ke ZHOU Yu-bin WANG Ai-lin XIAO Liang

中华医学杂志(英文版)2009,Vol.122Issue(2):140-144,5.
中华医学杂志(英文版)2009,Vol.122Issue(2):140-144,5.DOI:10.3760/cma.j.issn.0366-6999.2009.02.005

Clinical study of domestic ZQL-type retrievable vena cava filter

Clinical study of domestic ZQL-type retrievable vena cava filter

XU Ke 1ZHOU Yu-bin 2WANG Ai-lin 3XIAO Liang1

作者信息

  • 1. Department of Radiology, First Affiliated Hospital of China Medical University, Shenyang, Liaoning 110001, China
  • 2. Department of Vascular Surgery, People's Hospital of Liaoning Province, Shenyang, Liaoning 110016, China
  • 3. Department of Peripheral Vascular Disease, Heilongjiang Provincial Hospital, Haerbin, Heilongjiang 150036, China
  • 折叠

摘要

Abstract

Background Pulmonary thromboembolism (PTE) is a serious disease often leading to disability and death. Percutaneous placement of an inferior vena cava (IVC) filter is an effective method to prevent fatal PTE caused by lower extremity deep venous thrombosis (LEDVT). We developed the ZQL-type retrievable vena cava filter. The aim of this study was to evaluate the safety and effectiveness of the ZQL-type filter for prevention of fatal PTE. Methods A total of 144 patients with indications for placement of an IVC filter received insertion of filters via a femoral (n=37) or jugular (n=107) vein approach. Abdominal X-ray and color Doppler ultrasonography of IVC were regularly performed to visualize the position and condition of the IVC filter following filter placement. If thrombi in the lower extremity deep veins were removed within 2 weeks, the filter retrieval procedure was performed after an abdominal radiograph, an inferior vena cavogram and a pulmonary angiography. Otherwise, the filter should be kept permanently in place with regular follow-up inspections. Results One hundred and forty-four filters were implanted at the target sites with a success rate of 98.61% for one-time placement and 100% for two-time placement. Of the 137 patients followed up (follow-up rate, 95.14%), 43 patients had filters retrieved during a period of between 7 and 14 days (median, 12 days) following filter placement. The remaining 94 patients were followed for 7 days to 39 months (median follow-up period, 17 months). No filter migration or tilt, filter fracture or IVC perforation was observed. No obstruction of IVC occurred. No symptomatic PTE developed during the follow-up period. Conclusion The ZQL-type retrievable vena cava filter is a safe and effective device to prevent PTE. This filter possesses a distinctive stent-shape, stable design, high capture efficacy, and is easy to insert and retrieve and thus is suitable for clinical application.

关键词

pulmonary embolism/vena cava filters/venous thrombosis

Key words

pulmonary embolism/vena cava filters/venous thrombosis

分类

医药卫生

引用本文复制引用

XU Ke,ZHOU Yu-bin,WANG Ai-lin,XIAO Liang..Clinical study of domestic ZQL-type retrievable vena cava filter[J].中华医学杂志(英文版),2009,122(2):140-144,5.

基金项目

This project was supported by the grants from National Science and Technology Research Program in "The 9th Five-Year Plan"(No.96-907-01-02)and Liaoning Province Science and Technolgoy Commission(No.2003-362). (No.96-907-01-02)