In this paper a planar titanium oxide electrode was fabricated by standard microelectronic technol-ogy and its electrochemical properties in phosphate buffer solution was investigated. According to the cy-clic vohammetric results, both hydrogen peroxide and oxygen can be cathodic reduced at titanium oxide e-lectrode, while the reduction of hydrogen peroxide is preferential. The amperometric measurement of hy-drogen peroxide at the planar titanium oxide electrode in air saturated phosphate buffer solution at -300 mV shows fast responses. The amperometric response to 0.1 mmol/L hydrogen peroxide is 0.4 μA while the residual current due to the reduction of oxygen is only 14 nA. Glucose oxidase has been immobilized on the surface of titanium oxide film to explore the possible applications of this electrode in biosensing. The experimental results show it is difficult to immobilize enzyme on titanium oxide surface.关键词
二氧化钛电极/过氧化氢/溅射/还原Key words
titanium oxide/hydrogen peroxide/sputtering/reduction分类