眼科学报2002,Vol.18Issue(3):165-168,4.
Clinical Application of the Sapphire Unfolder Lens Injection System
Clinical Application of the Sapphire Unfolder Lens Injection System
Weiai Guo 1Danying Zheng 1Zhenyu Li 1Yiyong Qian 1Zhenping Zhang1
作者信息
- 1. Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou 510060, China
- 折叠
摘要
Abstract
Purpose: To summarize the clinical experience of 300 cases using the Sapphire unfloder intraocular lens (IOL) injection system.Methods: After the standard phacoemulsification, an AR40e IOL was implanted using the Sapphire Unfolder. The involved problems during and after the operation were observed and analyzed.Results:The complications occurred during the operation including the crack at the haptic-optic junction in 2 cases, slight kink in the haptic in 5 cases, IOL clamp into the cartridge in 2 cases, posterior capsular rupture in 2 cases and endothelium damage in the central small area in 4 cases. All the patients recovered successfully with IOLs in good position.Conclusion: IOL implantation with the Sapphire Unfolder led to no serious complications and got the satisfactory results.关键词
Unfolder injection system/foldable lens/IOL implantationKey words
Unfolder injection system/foldable lens/IOL implantation分类
医药卫生引用本文复制引用
Weiai Guo,Danying Zheng,Zhenyu Li,Yiyong Qian,Zhenping Zhang..Clinical Application of the Sapphire Unfolder Lens Injection System[J].眼科学报,2002,18(3):165-168,4.