The Influence of Central Corneal Thickness and Corneal Curvature and Axial Length on the Measurement of Intraocular Pressure
The Influence of Central Corneal Thickness and Corneal Curvature and Axial Length on the Measurement of Intraocular Pressure
Qing Li 1Minru Li 1Zhigang Fan 2Ningli Wang2
作者信息
- 1. Medical College of Sun Yat-sen University, Guangzhou 510089, China
- 2. Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou 510060, China .
- 折叠
摘要
Abstract
Purpose: To examine the influence of central corneal thickness (CCT), corneal curvature (CC), and axial length (AL) on intraocular pressure (IOP).Methods: Eighty-one clinically normal eyes were included in our study. The IOP, CCT, CC, AL were measured using a Goldmann applanation tonometer, optical pachymeter, keratometer and A-scan ultrasound biometer respectively in all subjects.Results: A highly significant positive correlation was identified between IOP and CCT. Linear regression analysis suggests that an increase in CCT of 0. 010 mm is associated with a 4. 946 mmHg increment in IOP. No significant positive correlation was identified between IOP and CC. P values are 0. 724 and 0.414 respectively for vertical and horizontal readings. A paradoxically reversed correlation was present between IOP and axial length.Conclusion: Corneal thickness is a very important confounding factor in the measurement of intraocular pressure, which warrants further attention in our clinical practice. 关键词
intraocular pressure/central corneal thickness/corneal curvature/axial lengthKey words
intraocular pressure/central corneal thickness/corneal curvature/axial length分类
医药卫生引用本文复制引用
Qing Li,Minru Li,Zhigang Fan,Ningli Wang..The Influence of Central Corneal Thickness and Corneal Curvature and Axial Length on the Measurement of Intraocular Pressure[J].眼科学报,2002,18(3):176-180,5.