组合式外固定器治疗股骨干稳定性骨折的生物力学研究
Biomechanical evolution of the combinable external fixtor for the treatment of stable femoral shaft fracture
刘志成 1赵峰 2魏若卉 1刘艳 1张涛1
作者信息
- 1. 首都医科大学生物医学工程学院 ,首都医科大学生物力学研究中心,
- 2. 中国科学院力学研究所微重力室,
- 折叠
摘要
Abstract
Four configurations of Combinable External Fixation(CEF) devices used for curing of fracture femur with the fracture gap cross the middle of the femur were tested to determine their biomechanical feature, such as the strain distribution of the femur-Fixator compound systems, the fixator stiffness. There were some differences between configurations of CEF in all modes of loading with respect to strain-distribution and stiffness. It is proved that the stiffness of the femur with CEF is stronger than that of intact femur. And the strain-distribution of the femur with CEF is also smoother than that of femur without CEF. In the stable fixation, because the load still go through the diaphysis, the CEF has a little stress shielding. And so the CEF can provide a better condition of the new bone formation. These data may be useful in developing the CEF for a particular clinical situation.关键词