Quantitative Measurements of the Decentration Distance of Four Kinds of Intraocular Lens
Quantitative Measurements of the Decentration Distance of Four Kinds of Intraocular Lens
Bo Feng 1Zhende Lin 1Jianjing Li 1Yonghua Li1
作者信息
- 1. Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou 510060, China
- 折叠
摘要
Abstract
Purpose:To investigate a quantitative method to measure decentration distance(DD) and analyze the DD of four kinds of intraocular lenses (IOLs).Methods :The pseudophakic eye photos were taken under the slit lamp after the IOLs were implanted a year or longer. Then the images were inputted into a computer and the Photoshop software was applied to analyze the DD. Four kinds of IOLs were analyzed,including AcrySof MA60BM (11 eyes), Allergan SA40N (11 eyes), AcrySof SA60AT/SA30AL (5 eyes), Corneal Quattro (4 eyes).Results:The mean value of DD was (0.264 3±0.157 4)mm in the AcrySof MA60BM group,(0.353 6±0.171 9)mm in the Array SA40N group,(0.309 9±0.152 9)mm in the SA60AT/SA30AL group, and (0.326 1±0.187 7)mm in the Quattro group, The difference of the DD values among these 4 kinds of IOLs had no statistic significance (F=0.506,P=0.681) and the values were not equal to zero (t=10.508, P < 0.001).
Conclusion:The Photoshop method is a simple and efficient way for DD analysis. The modern foldable posterior-chamber-IOLs have good centration performances when implanted into the capsular bags. Eye Science 2004;20:48-51.关键词
lens/intraocular decentration/quantitative measurementKey words
lens/intraocular decentration/quantitative measurement引用本文复制引用
Bo Feng,Zhende Lin,Jianjing Li,Yonghua Li..Quantitative Measurements of the Decentration Distance of Four Kinds of Intraocular Lens[J].眼科学报,2004,20(1):48-51,4.