Coriolis质量流量计中的相位差检测方法研究

仝猛 陈明 黄平云 任萍 梁彦 范东远 陈秋

测控技术2001,Vol.20Issue(1):8-10,3.
测控技术2001,Vol.20Issue(1):8-10,3.

Coriolis质量流量计中的相位差检测方法研究

Research on the Phase-difference Detection Method of Coriolis Mass Flowmeters

仝猛 1陈明 1黄平云 1任萍 1梁彦 1范东远 1陈秋1

作者信息

  • 1. 西北工业大学 测量与控制技术研究所,
  • 折叠

摘要

Abstract

Coriolis mass flowmeter is a precision flowmeter, which is able to handle an extensive range of applications and fluids, from fluids to gases.The mass flowrate is proportional to the phase difference of the two vibration signals detected from the vibrating tube of the meter,which fluid is passed through.A cross correlation method is presented to improve the phase detection precision,and thoroughly analyzes the laws between cross correlation and phase shift in theory.Then use Mathematica3.0 to simulate the cross correlation filtering of two vibration signals and compare the effects of the phase detection precision improvements.

关键词

互相关/傅立叶变换/相位差/流量计/信号处理

分类

机械制造

引用本文复制引用

仝猛,陈明,黄平云,任萍,梁彦,范东远,陈秋..Coriolis质量流量计中的相位差检测方法研究[J].测控技术,2001,20(1):8-10,3.