| 注册
首页|期刊导航|生命科学仪器|Data processing in pulse transit time measurement

Data processing in pulse transit time measurement

Jingmeng Zhang Ye Li

生命科学仪器2008,Vol.6Issue(10):28-32,5.
生命科学仪器2008,Vol.6Issue(10):28-32,5.

Data processing in pulse transit time measurement

Data processing in pulse transit time measurement

Jingmeng Zhang 1Ye Li2

作者信息

  • 1. Department of Biomedical Engineering,Xi'an Jiaotong University
  • 2. Institute of Biomedical and Health Engineering,SIAT,Shenzhen
  • 折叠

摘要

Abstract

Pulse transit time (PTT) is used as a noninvasive and cuff-less parameter to estimate blood pressure.In this paper,we develop an algorithm to ohtain PTT rapidly,which is appropriate for micro-processor and could achieve good accuracy in PTT,even in noisy measurements.The algorithm is based on finite impulse response(FIR) filter to reduce the noise and an adaptive threshold to detect the significant points of ECG and PPG.Evaluation of this method is hased on the signals from our PTT-based blood pressure devices It is shown that the method works well for PPT calculation.

关键词

transit time/data proeess/PPG/ECG

Key words

transit time/data proeess/PPG/ECG

引用本文复制引用

Jingmeng Zhang,Ye Li..Data processing in pulse transit time measurement[J].生命科学仪器,2008,6(10):28-32,5.