Abstract
Objective To explore a rapid method of calculating an percentage depth dose for irregular beam fields. Methods Worked out the percentage depth dose of irregular fields by means of A, B, C, D.Then compared the prescription doses and those measured. Results The average error of the method A is 3.67%,the average error of the method B is 0.47%.For the method C.the average error of rules block radiation beam fields is 0.24%, the average error of irregular block radiation beam fields is 0.69%.The average error of the method D is 0.28%. Conclusion In calculation of the percentage depth dose of iregular beam fields.the average error of the method B is 0.47%,it is less than ±l%.The method B is convenient for standardized operations and meets the accuracy demand.关键词
直线加速器/不规则照射野/百分深度剂量/放射治疗分类
医药卫生引用本文复制引用
徐宜武,林益匡..不规则射野的百分深度剂量计算[J].中国医疗设备,2011,26(2):39-41,3.