| 注册
首页|期刊导航|中国医疗设备|生化检验线性评价与应用

生化检验线性评价与应用

王淑媛 柯培锋 司徒瑞儒 姚珍 何卓雄 庾凤娟

中国医疗设备2011,Vol.26Issue(2):96-98,104,4.
中国医疗设备2011,Vol.26Issue(2):96-98,104,4.DOI:10.3969/j.issn.1674-1633.2011.02.031

生化检验线性评价与应用

Evaluation and Application of Clinical Biochemistry Examination Linearity

王淑媛 1柯培锋 2司徒瑞儒 1姚珍 1何卓雄 1庾凤娟1

作者信息

  • 1. 广州市海珠区第一人民医院,检验科,广东广州,510220
  • 2. 广东省中医院,检验科,广东广州,510220
  • 折叠

摘要

Abstract

Objective To explore the difference and adaptivity of linearity evaluation with three methods. Methods To evaluate the linearity of 25 clinical chemistry items with CAP-IRC, EP6-A and modified Doumas method. Results In EP6-A method, 17 of 25 items were evaluated direct linearity, 5 items were clinical acceptable linear, 3 items were non-linear. In CAP-IRC method,9 of 25 items were linear, 16 items were clinical acceptable linear. In modified Doumas method,15 of 25 items were linear, 10 items were non linear. Conclusion EP6-A linearity evaluation method is more objective and more suitable to clinical biochemistry.

关键词

CAP-IRC/EP6-A/改良的Doumas方法/线性评价/生化分析仪

分类

数理科学

引用本文复制引用

王淑媛,柯培锋,司徒瑞儒,姚珍,何卓雄,庾凤娟..生化检验线性评价与应用[J].中国医疗设备,2011,26(2):96-98,104,4.

中国医疗设备

OACSTPCD

1674-1633

访问量0
|
下载量0
段落导航相关论文