医疗卫生装备2011,Vol.32Issue(5):98-101,4.
心电图机ADS功能异常的原因分析
Probable Causation of Abnormal Phenomenon of Electrocardiograph ADS Functionality
摘要
Abstract
Objective To research the probable causation of abnormal phenomemom of electrocardiograph ADS functionality.Methods The probable causation was analyzed from digital filter method, time constant, frequency superposition,and so forth. Reaults digital filter method and less time constant were probable causations of low frequency signal distortion, while frequency superposition was the probable causation of high frequency signal drifting. Conclusion Digital filter method,less time constant and frequency superposition are the probable causations of the abnormal phenomenon of electrocardiograph ADS functionality.[Chinese Medical Equipment Journal , 2011 ,32( 5 ) : 98-101]关键词
抗漂移系统/低频信号/失真/高频信号/缓慢飘移Key words
anti-drifting system (AOS)/ low frequency signal/ distortion/ high frequency signal/ driftiing分类
机械制造引用本文复制引用
高原,周娟,刘光荣..心电图机ADS功能异常的原因分析[J].医疗卫生装备,2011,32(5):98-101,4.基金项目
国家自然基金课题(60801009,60971044) (60801009,60971044)