国际医药卫生导报2011,Vol.17Issue(11):1376-1378,3.DOI:10.3760/cma.j.issn.1007-1245.2011.11.034
Roche Modular PPI全自动生化分析仪测定Glu的性能评价
Performance of Roche Modular PPI automatic biochemical analyzer for glucose detection
李倩珺
1吴晓宾 2柯培锋2
作者信息
- 1. 510000,广州医学院第一附属医院检验科
- 2. 510120,广州,广东省中医院检验科
- 折叠
摘要
Abstract
Objective To evaluate the performance of Roche Modular PPI automatic biochemical analyzer for glucose detection. Methods Normal or pathological level of serum in the same batches was detected 20 times within a day and the within-run %CV was calculated; then the serum was detected once a day for 20 days and the between-run %CV was calculated. According to the date on external quality assessment from the Ministry of Pubic Health in 2010 and the within-run %CV and between-run %CV, the extend inaccuracy was calculated. The glucose level in the samples diluted on a specific ratio was detected, the linear range was verified. Results The within-run %CV was 0.49% for the normal level of quality control serum and 0.24% for the pathological level; the between-run %CV was 0.58% and 0.27% and the inaccuracy was ±3.1% and ± 2.2%, respectively. The linear range was 0 - 32mmol/L. Conclusions The performance of Roche Modular PPI automatic biochemical analyzer for glucose detection can satisfied the demand of clinical chemical labs.关键词
全自动生化分析仪/血糖/线性范围Key words
Automatic biochemical analyzer/Glucose/Linear range引用本文复制引用
李倩珺
,吴晓宾,柯培锋..Roche Modular PPI全自动生化分析仪测定Glu的性能评价[J].国际医药卫生导报,2011,17(11):1376-1378,3.