原子能科学技术2011,Vol.45Issue(3):360-364,5.
新型PAGAT聚合物凝胶剂量计的磁共振成像研究
Magnetic Resonance Imaging Stud of New PAGAT Polymer Gel Dosimeter
摘要
Abstract
A new type of PAGAT polymer gel dosimeter, containing acrylamide,gelatin, N, N'-methylene-bisacrylamide (BIS), tetrakis (hydroxymethyl) phosphonium chloride (THPC) and formaldehyde, was prepared under normal atmospheric conditions. The results show that its transverse relaxation ratio increases with the absorbed dose and its minimum detection limit is given. The dosimeter has good uniformity of one batch (better than 3. 2%) and negligible dose-rate dependence in the range of 0. 051.00 Gy/s. Using the gel dosimeter, the dose distribution of a heterogeneous interface was obtained and the dose enhancement effect was analyzed.关键词
聚合物凝胶剂量计/PAGAT/磁共振成像/放射治疗/剂量验证分类
通用工业技术引用本文复制引用
赵云,王飞,何承发,刘艳,杨进军,卫平强,兰博,崔江维,费武雄,李茂顺..新型PAGAT聚合物凝胶剂量计的磁共振成像研究[J].原子能科学技术,2011,45(3):360-364,5.基金项目
国家自然科学基金资助项目(10775178) (10775178)