首页|期刊导航|中华医学杂志(英文版)|Reproducibility and accuracy of quantitative assessment of articular cartilage volume measurements with 3.0 tesla magnetic resonance imaging

Reproducibility and accuracy of quantitative assessment of articular cartilage volume measurements with 3.0 tesla magnetic resonance imaging

XING Wei SHENG Jing CHEN Wen-hua TIAN Jian-ming ZHANG Li-rong WANG Dong-qing

中华医学杂志(英文版)2011,Vol.124Issue(8):1251-1256,6.
中华医学杂志(英文版)2011,Vol.124Issue(8):1251-1256,6.DOI:10.3760/cma.j.issn.0366-6999.2011.08.024

Reproducibility and accuracy of quantitative assessment of articular cartilage volume measurements with 3.0 tesla magnetic resonance imaging

Reproducibility and accuracy of quantitative assessment of articular cartilage volume measurements with 3.0 tesla magnetic resonance imaging

XING Wei 1SHENG Jing 1CHEN Wen-hua 1TIAN Jian-ming 1ZHANG Li-rong 1WANG Dong-qing1

作者信息

  • 折叠

摘要

Abstract

Background Quantitative magnetic resonance imaging (qMRI) of articular cartilage represents a powerful tool in osteoarthritis research, but has so far been confined to a field strength of 1.5 T. The aim of the study was to determine the reproducibility and accuracy of qMRI assessments of the knee cartilage volume by comparing quantitative swine cartilage volumes of the sagittal (sag) multi echo data imagine combination water-excitation (MEDICwe) sequence and the fast low-angle shoot water-excitation (FLASHwe) sequence at 3.0-T MRI to directly measured volumes (DMV) of the surgically removed articular cartilage.Methods Test-retest MRI was acquired in 20 swine knees. Two sag FLASHwe sequences and two sag MEDICwe sequences (spatial resolution 0.4 mm × 0.4 mm × 1.0 mm of 3-dimension (3D) were acquired at 3-T MRI in a knee.Articular cartilage volume was calculated from 3D reformations of the MRI by using a manual program. Calculated volumes were compared with DMV of the surgically removed articular cartilage. Knee joint cartilage plates were quantified paired in order.Results In the knee joint of swine, reproducibility errors (paired analysis) for cartilage volume were 2.5% to 3.2% with sag FLASHwe, and 1.6% to 3.0% with sag MEDICwe. Correlation coefficients between results obtained with qMRI and DMV ranged from 0.90 to 0.98 for cartilage volume. Systematic pairwise difference between results obtained with qMRI and DMV ranged from -1.1% to 2.8%. Random pairwise differences between results obtained with qMRI and DMV ranged from (2.9 ±2.4)% to (6.8±4.5)%.Conclusions FLASHwe and MEDICwe sequences permit highly accurate and reproducible analysis of cartilage volume in the knee joints of swine at 3-T MRI. Cartilage volume reproducibility for the MEDICwe data is slightly higher than the FLASHwe data.

关键词

accuracy/cartilage/magnetic resonance imaging/reproducibility/volume

Key words

accuracy/cartilage/magnetic resonance imaging/reproducibility/volume

引用本文复制引用

XING Wei,SHENG Jing,CHEN Wen-hua,TIAN Jian-ming,ZHANG Li-rong ,WANG Dong-qing..Reproducibility and accuracy of quantitative assessment of articular cartilage volume measurements with 3.0 tesla magnetic resonance imaging[J].中华医学杂志(英文版),2011,124(8):1251-1256,6.

基金项目

This work was supported by a grant from Shanghai Science and Technology Committee (No.08210707600) (No.08210707600)