Availability of solid phantom for accelerator absorbed dose measurement

陈伟思 1李国庆 1林意群 1邓晓刚 2许志新2

作者信息

  • 1. 510515,广州,南方医科大学生物医学工程学院
  • 2. 510515,广州,南方医院放疗科
  • 折叠

摘要

Abstract

Objective To verify the feasibility and reliability of using solid phantom for absorbed dose measurement and analyze the measurement result. Methods The dosimeter was used to measure the absorbed dose of Varian23EX medical linear accelerator in solid phantom and standard water phantom at the depth of calibration,and the correction factor of solid phantom versus standard value was given. Results The ratio of absorbed dose measured by home-made solid phantom and standard water phantom was betwcen 0.980-1.010. Conclusion The ABS resin home-made solid phantom, which is easily used and less complicated to operate and with a good repeatability, can be used to measure absorbed dose with the use of correction factor for the measurement correction.[Chinese Medical Equipment Journal , 2011 , 32 ( 4) : 16-18]