生态学杂志2011,Vol.30Issue(5):939-943,5.
贵阳市大气气态总汞:Lumex RA-915AM与Tekran2537A的对比观测水
Atmospheric total gaseous mercury concentration in Guiyang: Measurements intercomparison with Lumex RA-915AM and Tekran 2537A.
摘要
Abstract
From May 12 to June 19 , 2010, a high-time resolution observation on the atmospheric total gaseous mercury (TGM) concentration in an urban area of Guiyang was continuously conducted by using Lumex RA-915AM ( made in Russia) and Tekran 2537A ( made in Canada). Though the two sampling instruments have distinct operation principles, the overall characteristics and temporal distribution patterns of measured atmospheric TGM concentration were quite consistent, indicating that both of the instruments could accurately measure the atmospheric TGM concentration. The atmospheric TGM concentration in Guiyang was much higher than the mean global background value, and also, higher than the atmospheric TGM concentration in urban areas of Europe and North America. The preliminary sources identification suggested that the point source mercury emission from domestic activities played an important role in affecting the atmospheric TGM concentration in Guiyang, and some large factories located at the east of Guiyang were also the important sources to the regional atmospheric mercury.关键词
大气气态总汞/贵阳市/污染源/方法对比Key words
atmospheric total gaseous mercury/ Guiyang/ emission source/ methods intercomparison .分类
农业科技引用本文复制引用
付学吾,冯新斌,张辉..贵阳市大气气态总汞:Lumex RA-915AM与Tekran2537A的对比观测水[J].生态学杂志,2011,30(5):939-943,5.基金项目
国家自然科学基金项目(40773067和41003051)和贵州省2009年度科学技术基金项目(黔科合J字[2009]2257号)资助. (40773067和41003051)