Abstract
Objective To study the uniformity and influential factors in detecting RBC or WBC by urine dipstick analysis and UF1000i urinary sediment analyzer. Methods Fresh unne samples from 1000 patients were detected for RBC and WBC by urine dipstick analysis, UF1000i and DiaSys R/S 2003 urinary sediment analyzer. The uniformity and influential factors were compared and examined. Results The positive coincidence rate of UF1000i and urine dipstick analysis for RBC and WBC detection was higher than that of DiaSys R/S 2003 and urine dipstick analysis, and such positive coincidence rate was effected by the amount of RBC and WBC. Conclusion UF1000i urinary sediment analyzer was superior to DiaSys R/S 2003 urinary sediment analyzer. It was the more ideal test platform for urine RBC and WBC detection.[Chinese Medical Equipment Journal, 2011 , 32 (7) : 61-62]关键词
UF1000i/DiaSys/R/S 2003/尿沉渣定量Key words
UF1000i/ DiaSys R/S 2003/ urinary sediment count分类
医药卫生引用本文复制引用
杜秀敏,孙子涵,孔立新..尿干化学法联合UF1000i、DiaSys R/S2003尿沉渣定量检测尿红、白细胞的一致性比较[J].医疗卫生装备,2011,32(7):61-62,2.