ACL TOP全自动血凝仪抗干扰能力评价

陆学军 张将

中国医疗设备2011,Vol.26Issue(8):96-100,5.
中国医疗设备2011,Vol.26Issue(8):96-100,5.DOI:10.3969/j.issn.1674-1633.2011.08.035

ACL TOP全自动血凝仪抗干扰能力评价

The Evaluation of Anti-interference Ability of ACL TOP

陆学军 1张将2

作者信息

  • 1. 卫生部北京医院,检验科,北京,100730
  • 2. 中南大学湘雅医学院,检验系,湖南,长沙,410013
  • 折叠

摘要

Abstract

Objective To study the anti-interference ability of the ACL TOP hemostasis analyzer. Methods PT,AlyTT,TT and FIB were detected twenty times respectively in normal appearance plasma. The different concentrationa of interferents (hemolysiB.lipemia.chylous samples) were put respectively into experimental plasmas ,in which PT.APTT,TT and FIB were measured. Results When PT, APTT,TT and FIB were measured repeatedly for precision test,CV was less than 4 %.For accuracy test P>0.05. The ability of anti-hemolytic. jaundice interference is good, but high concentrations of triglycerides interfered with the test results of PT and APTT. Conclusion The ACL TOP coagulation analyzer,with a good precision and accuracy result. has a great anti-interference ability of bemolysis and jaundice samples. which also has a certain ability of anti-TG interference.

关键词

血凝分析仪/血细胞分析仪/生化分析仪/抗干扰能力评价

Key words

coagulation analyzer/ blood cell analyzer/ biochemical analyzer/ evaluation of anti-interference ahility

分类

医药卫生