中国医疗设备2011,Vol.26Issue(8):96-100,5.DOI:10.3969/j.issn.1674-1633.2011.08.035
ACL TOP全自动血凝仪抗干扰能力评价
The Evaluation of Anti-interference Ability of ACL TOP
陆学军 1张将2
作者信息
- 1. 卫生部北京医院,检验科,北京,100730
- 2. 中南大学湘雅医学院,检验系,湖南,长沙,410013
- 折叠
摘要
Abstract
Objective To study the anti-interference ability of the ACL TOP hemostasis analyzer. Methods
PT,AlyTT,TT and FIB were detected twenty times respectively in normal appearance plasma. The different
concentrationa of interferents (hemolysiB.lipemia.chylous samples) were put respectively into experimental
plasmas ,in which PT.APTT,TT and FIB were measured. Results When PT, APTT,TT and FIB were
measured repeatedly for precision test,CV was less than 4 %.For accuracy test P>0.05. The ability of
anti-hemolytic. jaundice interference is good, but high concentrations of triglycerides interfered with the
test results of PT and APTT. Conclusion The ACL TOP coagulation analyzer,with a good precision and
accuracy result. has a great anti-interference ability of bemolysis and jaundice samples. which also has a
certain ability of anti-TG interference.关键词
血凝分析仪/血细胞分析仪/生化分析仪/抗干扰能力评价Key words
coagulation analyzer/ blood cell analyzer/ biochemical analyzer/ evaluation of anti-interference ahility分类
医药卫生引用本文复制引用
陆学军,张将..ACL TOP全自动血凝仪抗干扰能力评价[J].中国医疗设备,2011,26(8):96-100,5.