祁增凌 钟姗姗 苏韶生

中国医学装备2011,Vol.8Issue(7):82-84,3.
中国医学装备2011,Vol.8Issue(7):82-84,3.

GE Senographe2000D数字化乳腺机的维修

Troubleshooting for GE Senographe 2000D digital breast X-ray machine

祁增凌 1钟姗姗 1苏韶生1

作者信息

  • 1. 中山市人民医院临床医学工程中心,广东,中山,528403
  • 折叠

摘要

Abstract

Objective: To investigate the troubleshooting of the GE Senographe 2000D digital breast X-ray machine. Methods: Error codes of the GE Senographe 2000D can help user to locate the failure and learn more about the failure description. The user can analyze the hardware failure through consulting the circuit diagram and software failure can be resolved by reinstall the software. Results: The compression failure has been successfully removed. Conclusion: The failure can be resolved by oneself once the user can proficiently operate the machine correctly and know how to install the software. Meanwhile, it requires the user to understand the maintance ideas and analyze the circuit diagram correctly.

关键词

乳腺机/Senographe/维修

Key words

Breast equipment/ Senographe/ Maintenance

分类

医药卫生

引用本文复制引用