Objective To explore the differences between the values of blood pressure at human brachial artery and canine femoral artery taken by mercurial sphygmomanometer and electronic sphygmomanometer. Methods The study involved 340 patients and 20 healthy dogs which were taken blood pressure by both mercurial sphygmomanometer and electronic sphygmomanometer for the values of blood pressure at brachial artery and canine femoral artery and then the comparison was conducted about the values. Results There was no significant difference between the values of blood pressure taken by mercurial sphygmomanometer and electronic sphygmomanometer in patients and in dogs (P>0.05). Conclusion It is accurate to take blood pressure value by electronic sphygmomanometer.