基于体模的肺内磨玻璃密度结节容积测量准确性研究

鄂林宁 李健丁 武志峰

山西医科大学学报2011,Vol.42Issue(8):655-658,4.
山西医科大学学报2011,Vol.42Issue(8):655-658,4.DOI:10.3969/J.ISSN.1007-6611.2011.08.012

基于体模的肺内磨玻璃密度结节容积测量准确性研究

The accuracy of volumetric measurement of pulmonary ground-glass opacity nodules based on the chest phantom

鄂林宁 1李健丁 1武志峰1

作者信息

  • 1. 山西医科大学第一临床医学院放射科,太原,030001
  • 折叠

摘要

Abstract

Objective To evaluate the effect of various tube currents on accuracy of volumetric measurement of ground-glass opacity nodules. Methods A chest phantom and 13 artificial ground-glass opacity nodules with known volume were scanned using a 64-sec-tion computed tomography ( CT ) scanner at different tube currents ( 30,60,90,120,150,180,210 mA ). All data were reconstructed with slice thickness of 0. 625 mm and bone algorithm. Automated nodule volume measurements were performed using computer-assisted volume measurement software. Nodule volumes measured by software were compared with the reference-standard volume of ground-glass opacity nodules. Correlation coefficient was used for statistical analysis. The relative percent error ( RPE ) between the CT measured volume and the reference-standard volume of ground-glass opacity nodules was measured. Results All volumes measured by MSCT at different tube currents were correlated with the reference-standard volume. The volumes of ground-glass opacity nodules measured at 30, 60,90 mA were underestimated compared with the reference standard volume,and their mean RPEs were ( -5. 63 ± 8. 60 )% ,( -6. 07 ±7.49 )% ,( -5.50 ±8.31% ),respectively. While there was substantial overestimation of volume at 120,150,180 and 210 mA, and the mean RPEs were ( 2. 32 ± 8. 05 )% ,( 3. 06 ± 8. 18 )% ,( 4. 71 ±7. 63 )% ,( 3. 05 ±7. 79 )% .respectively. Conclusion Computer-assisted volume measurement is a promising method for quantification of ground-glass opacity nodule volume. And the different tube currents could have effects on the accuracy of volumetric measurement.

关键词

体层摄影术/X线计算机/体积测量/磨玻璃密度结节

Key words

tomography,X-ray computed/volume measurement/ground-glass opacity nodule

分类

医药卫生

引用本文复制引用

鄂林宁,李健丁,武志峰..基于体模的肺内磨玻璃密度结节容积测量准确性研究[J].山西医科大学学报,2011,42(8):655-658,4.

基金项目

山西省科技攻关项目(20090311057-3) (20090311057-3)

山西医科大学学报

OACSTPCD

1007-6611

访问量1
|
下载量0
段落导航相关论文