辐射防护2012,Vol.32Issue(1):53-59,64,8.
~(222)Rn/~(220)Rn子体测量方法与仪器比较
Comparison of Methods and Instruments for ~(222)Rn/~(220)Rn Progeny Measurement
刘艳阳 1尚兵 2周青芝 3武云云2
作者信息
- 1. 中国原子能科学研究院,北京102413
- 2. 中国疾病预防控制中心辐射防护与核安全医学所,北京100088
- 3. 南华大学核工程与核技术学院,湖南衡阳421001
- 折叠
摘要
Abstract
In this paper,comparisons were made among three methods of measurement(grab measurement,continuous measurement and integrating measurement) and also measurements of different instruments for a radon/thoron mixed chamber.Taking the optimized five-segment method as a comparison criterion,for the equilibrium-equivalent concentration of 222 Rn,measured results of BWLM and 24 h integrating detectors are 31% and 29% higher than the criterion,the results of Wlx, however,is 20% lower;and for 220 Rn progeny,the results of FJ414-142,KF-602D,BWLM and 24 h integrating detector are 86%,18%,28% and 36% higher than the criterion respectively,except that of WLx,which is 5% lower.For the differences shown,further research is needed.关键词
222Rn子体/220Rn子体/抓取测量/连续测量/累积测量Key words
222 Rn progeny/220 Rn progeny/grab sample measurement/continuous measurement/integrating measurement分类
资源环境引用本文复制引用
刘艳阳,尚兵,周青芝,武云云..~(222)Rn/~(220)Rn子体测量方法与仪器比较[J].辐射防护,2012,32(1):53-59,64,8.