Abstract
This study adopts Flow Injection Analysis for online determination of the Anionic Surfaetant in water. The autoanalyzer made by Bran + Luebbe corporation in Germany was applied for determination. In the range of 0. 050 - 0. 400 mg/L, it is available to get good linear equations with correlation coefficient 〉 0, 9990, detection limit of 0. 015 mg/L and water samples recovery of 92% - 109%. This method meets the requirement of determining anionic surfactant for water quality control. The factors affecting the determination of anionic surfactant are discussed as well.关键词
流动注射/阴离子表面活性剂/水质/影响因素Key words
Flow injection/anionic surfaetant/water quality/affecting factors分类
资源环境引用本文复制引用