| 注册
首页|期刊导航|四川环境|流动注射法测定水中阴离子表面活性剂的探讨

流动注射法测定水中阴离子表面活性剂的探讨

娄明华 施新锋

四川环境2012,Vol.31Issue(1):46-48,3.
四川环境2012,Vol.31Issue(1):46-48,3.

流动注射法测定水中阴离子表面活性剂的探讨

Discussion on Determination of Anionic Surfactant in Water Using Flow Injection Analysis

娄明华 1施新锋1

作者信息

  • 1. 无锡市环境监测中心站,江苏无锡214023
  • 折叠

摘要

Abstract

This study adopts Flow Injection Analysis for online determination of the Anionic Surfaetant in water. The autoanalyzer made by Bran + Luebbe corporation in Germany was applied for determination. In the range of 0. 050 - 0. 400 mg/L, it is available to get good linear equations with correlation coefficient 〉 0, 9990, detection limit of 0. 015 mg/L and water samples recovery of 92% - 109%. This method meets the requirement of determining anionic surfactant for water quality control. The factors affecting the determination of anionic surfactant are discussed as well.

关键词

流动注射/阴离子表面活性剂/水质/影响因素

Key words

Flow injection/anionic surfaetant/water quality/affecting factors