图说质谱仪TRITON离子源室真空问题的解决

吴桐开

现代科学仪器Issue(5):131-134,4.
现代科学仪器Issue(5):131-134,4.

图说质谱仪TRITON离子源室真空问题的解决

Illustration of How to Solve the Ion Source Chamber Vacuum Issues for the Mass Spectrometer TRITON

吴桐开1

作者信息

  • 1. 武汉地质矿产研究所,武汉430223
  • 折叠

摘要

Abstract

Thermal Ionization Mass Spectrometer TRITON has extremely high requirements on the ion source chamber.Once the chamber leaks,it will result in the instability of ion flow or even failure of high-voltage spark,etc.At the very least,it cannot collect data and the spectrometer cannot work properly,and it will result in the broken of the spectrometer in a clumsy way.This paper provides a detailed description of this usual phenomenon,and gives profound analysis of the failure causes.It confirms that the sealing performance deterioration of the drive axle of the sample wheel causes the leakage.In order to save costs and get the spectrometer resume its normal work as soon as possible,the author decided to disassemble and fix the spectrometer by himself and record the whole procedure with pictures.

关键词

热电离质谱仪/离子源室/真空

Key words

Thermal ionization mass spectrometer/Ion source chamber/Vacuum