医用直线加速器感生放射性冷却时间的检测

屠波 高关心 安文昊 卜羽 朱永丽 夏慧琳

中国医疗设备2011,Vol.26Issue(12):23-24,16,3.
中国医疗设备2011,Vol.26Issue(12):23-24,16,3.DOI:10.3969/j.issn.1674-1633.2011.12.005

医用直线加速器感生放射性冷却时间的检测

Study of Induced Radioactivity and Measure of Cooling Time to the Medical Linear Accelerator

屠波 1高关心 1安文昊 1卜羽 1朱永丽 1夏慧琳1

作者信息

  • 1. 内蒙古自治区人民医院医学工程处,内蒙古呼和浩特 010017
  • 折叠

摘要

Abstract

Objective Study the relationship between the induced radioactivity and the cooling time of medical linear accelerator by the detection of induced radioactivity. Methods When the X-ray energy of the accelerator were 6MV, 10 MV and 15MV, use radiation detection apparatus and 451P inspection to detect the induced radioactivity of 200cGy radiation dose within 0-30 minutes. Analyze and make statistical analysis test data. Results When the X-ray energy of accelerator increases, the induced radioactivity incr eases, and the cooling time is extended accordingly. Conclusion X-ray energy and cooling time affect the induced radioactivity levels of Medical linear accelerator.

关键词

医用直线加速器/感生放射性/冷却时间

Key words

medical linear accelerator/induced radioactivity/cooling time

分类

医药卫生