XT-4000i多功能全自动血细胞分析仪在体液细胞检测中与手工法的应用比较

白雪丽

中国医疗设备2012,Vol.27Issue(1):75-76,2.
中国医疗设备2012,Vol.27Issue(1):75-76,2.DOI:10.3969/j.issn.1674-1633.2012.01.024

XT-4000i多功能全自动血细胞分析仪在体液细胞检测中与手工法的应用比较

Application of XT-4000i Multi-function Automated Blood Cell Analyzer in Body Fluid Detection

白雪丽1

作者信息

  • 1. 山东大学附属省立医院检验科,山东济南250021
  • 折叠

摘要

Abstract

Objective To investigate the application of XT-4000i multi-function automated blood cell analyzer in body fluid detection. Methods 210 cases of body fluid were collected from inpatients and outpatients randomly, and were tested with the XT-4000i multi-function automated blood cell analyzer and the manual method, and then the correlation of the white blood cell count ,red blood cell count and differential count of the body fluid of the two methods were analyzed. Results Good agreement for the white blood cell count, red blood cell count and differential count were obtained between the XT-4000i multifunction automated blood cell analyzer and the manual method, and the correlation coefficient(r) of the white blood cell count, red blood cell count and differential count were 0.968,0.973 and 0.936 respectively. Conclusion the XT-4000i multi-function automated blood cell analyzer has high value in clinic body fluid detection.

关键词

XT-4000i多功能血细胞分析仪/体液细胞检测/白细胞计数/红细胞计数/白细胞分类

Key words

XT-4000i multi-function automated blood cell analyzer/ body fluid detection/ white blood cell count/ red blood cell count/ white blood cell differential count