| 注册
首页|期刊导航|医疗卫生装备|螺旋断层调强放疗治疗宫颈癌误差分析

螺旋断层调强放疗治疗宫颈癌误差分析

张慧娟 周桂霞 徐寿平 解传斌 戴相昆 鞠忠建

医疗卫生装备2012,Vol.33Issue(1):65-66,2.
医疗卫生装备2012,Vol.33Issue(1):65-66,2.

螺旋断层调强放疗治疗宫颈癌误差分析

Setup Error Analysis of Helical Tomotherapy for Cervix Cancer

张慧娟 1周桂霞 1徐寿平 1解传斌 1戴相昆 1鞠忠建1

作者信息

  • 1. 解放军总医院放疗科,北京100853
  • 折叠

摘要

Abstract

Objective To analyze inter-fractional setup errors during helical tomotherapy for cervix cancer using daily online MVCT IGRT technique. Methods From Jan.2009 to Dec.2010, 30 cervix cancer patients were treated by helical tomotherapy. Before treatment, all patients underwent MVCT scanning. MVCT images were registered with automatic KVCT plan and used to readjust the registration according to the dose distribution. The setup errors were obtained and analyzed. Results A total of 752 online MVCT scannings were performed for these 30 patients. The setup errors on X,Y,Z directions were (1.9+1.61)mm, (5.73+1.31)mm and (0.61+1.06)mm, respectively. Conclusion Daily online pre-treatment IGRT correction can reduce inter-fractional setup errors. MVCT can improve the accuracy and precisions of cervix cancerradiotherapy treatment.[Chinese Medical Equipment Journal,2011,33(1) :65-66]

关键词

宫颈癌/图像引导放疗/摆位误差/螺旋断层放疗

Key words

cervix cancer/ IGRT/ setup error/ tomotherapy

分类

医药卫生

引用本文复制引用

张慧娟,周桂霞,徐寿平,解传斌,戴相昆,鞠忠建..螺旋断层调强放疗治疗宫颈癌误差分析[J].医疗卫生装备,2012,33(1):65-66,2.

医疗卫生装备

OACSTPCD

1003-8868

访问量0
|
下载量0
段落导航相关论文