首页|期刊导航|医疗卫生装备|Beckman-Coulter LH750全自动血液分析仪校准

Beckman-Coulter LH750全自动血液分析仪校准

周腾坚

医疗卫生装备2012,Vol.33Issue(1):113-114,2.
医疗卫生装备2012,Vol.33Issue(1):113-114,2.

Beckman-Coulter LH750全自动血液分析仪校准

Calibration for Automatic Blood Analyzer of Beckman-Coulter LH750

周腾坚1

作者信息

  • 1. 永嘉县人民医院检验科,浙江永嘉325100
  • 折叠

摘要

Abstract

Objective To calibrate the automatic blood analyzer of Beckman-Coulter LH750. Methods With the environment requirements satisfied, the routine maintenance and special maintenance, including incandesce of counting hole, flow line cleaning and so on, were performed. The analyzer was powered off and then on after a two-hour soaking. After the tests for repeatability and residual contamination rate, the calibrator S-CAL(Lot:4754) provided by Beckman Coulter Company was used for calibration, then calibration coefficients for automatic and manual injection were obtained. Results The device proved to be gifted with high repeatability, low residual contamination rate, whose variation coefficient was below the lower limit of the defined value. The new calibration coefficient was helpful to enhance the accuracy of the device. Conclusion The automatic blood analyzer has to be calibrated periodically to improve the test quality. [Chinese Medical Equipment Journal,2011,33(1):113-114]

关键词

全自动血液分析仪/校准/校准系数/溯源性

Key words

automatic blood analyzer/ calibration/ calibration coefficient/ traceability