辐射防护2012,Vol.32Issue(2):103-107,5.
CR39中子个人剂量计性能实验研究
Study on Performance Test of CR39 Personal Dosimeter for Fast Neutron
曹磊 1邓君 1张贵英 2郭文 1钱叶侃 1胡爱英 1吉艳琴1
作者信息
- 1. 中国疾病预防控制中心辐射防护与核安全医学所,北京100088
- 2. 中国原子能科学研究院,北京102413
- 折叠
摘要
Abstract
Neutron personal dose monitoring is an important part of external monitoring for occupational exposure.According to the principles of neutron detecting of CR39 and its structure,experiments regarding energy response and angle response as well as dose equivalent response of CR39 neutron personal dosimeter against neutron flux emitted by Am-Be were conducted.Results show a strong dependence of neutron flux sensitivity on the energy and incident angle of neutron.The critical angle of 64° for incident neutron and neutron flux sensitivity of about 2.24×10-4 trs/n were obtained.关键词
CR39固体核径迹探测器/中子个人剂量监测/性能实验Key words
CR39 SSNTD/neutron personal dose monitoring/performance test分类
能源科技引用本文复制引用
曹磊,邓君,张贵英,郭文,钱叶侃,胡爱英,吉艳琴..CR39中子个人剂量计性能实验研究[J].辐射防护,2012,32(2):103-107,5.