首页|期刊导航|医疗卫生装备|应用OBI系统分析鼻咽癌调强放疗的摆位误差

应用OBI系统分析鼻咽癌调强放疗的摆位误差

叶森林 梁廷 荣知璧

医疗卫生装备2012,Vol.33Issue(3):64-65,2.
医疗卫生装备2012,Vol.33Issue(3):64-65,2.

应用OBI系统分析鼻咽癌调强放疗的摆位误差

Setup Error Analysis with on Board Imager System for Nasopharyngeal Carcinoma Treated by Intensity Modulated Radiotherapy

叶森林 1梁廷 1荣知璧1

作者信息

  • 1. 重庆医科大学附属第一医院肿瘤科放疗中心,重庆400016
  • 折叠

摘要

Abstract

Objective To analyze the setup errors using on board imager(OBI) system in nasopharyngeal carcinoma treated by intensity modulated radiotherapy. Methods Varian-21EX linear accelerator was used for medical treatment of 38 nasopharyngeal carcinoma cases. Cone-beam computed tomography(CBCT) images were taken by OBI system before the first intensity modulated radiotherapy, once a day for the first 3 days, once a week later, and the translation and rotation errors between were obtained by matching the CBCT images and the planning CT images. Results The average errors obtained from the first scan included right-left (-0.49±1.90) mm, head-feet (0.64+1.36) mm, anteriority-posterior (-0.57± 1.78) mm, and rotation (0.05±0.11) °. After the correction for the first time, the setup errors were all reduced, including right-left (0.30±1.01) mm, head-feet (0.12±0.92) mm, anteriority-posterior (0.47 ±1.05) mm, and rotation (-0.02±0.03) °. Conclusion The setup errors verified by using OBI system can improve the positioning accuracy, and reduce the irradiation amount for surrounding normal tissues. For the intensity modulated radiotherapy of nasopharyngeal carcinoma, the putting boundary is about 2mm from the clinical target volume(CTV) to the planning target volume(PTV). [Chinese Medical Equipment Journal,2011,33(3):64-65]

关键词

摆位误差/OBI/调强放射治疗

Key words

setup error/on board imager/intensity modulated radiotherapy

分类

医药卫生

引用本文复制引用

叶森林,梁廷,荣知璧..应用OBI系统分析鼻咽癌调强放疗的摆位误差[J].医疗卫生装备,2012,33(3):64-65,2.

医疗卫生装备

OACSTPCD

1003-8868

访问量0
|
下载量0
段落导航相关论文