Pentacam与IOLMaster测量前房深度的一致性和可重复性研究

Pentacam and IOLMaster in anterior chamber depth measurements:Agreement and Repeatability

杨君 1郝更生1

作者信息

  • 1. 614000 四川省乐山市人民医院眼科
  • 折叠

摘要

Abstract

Objective To explore the differences of anterior chamber depth (ACD) determined by Pentacam and IOLMaster and to evaluate die agreement of the two instruments. Methods ACD of 61 right eyes were measured-with Pen-tocam and IOLMaster. Paired t lest was used to analyze die difference of mean ACD between nSe two methods,and me Bland-Altman method was used to erahiate the agreement of die two methods, ICC was used to evaluated die repeatability. Kcsons The mean ACD nine was (3.57 ±0.34) mn for Pentacam and (3.54 ±0. 34) mm for IOLMasler ( P =0. 023). Tne Bland-Altmann sbowed that the tiro methods had Mimparable results. Condosfaa Pentacam and IOLMaster has agreement in ACD measurement, and bom method have good repeatability.

关键词

Pentacam/IOLMaster/前房深度/一致性/可重复性

Key words

Pentacam/IOLMaster/Anteriorchamentdep:Agreement/Repeatability