核化学与放射化学2012,Vol.34Issue(2):124-128,5.
光纤光度计测定铀及铀铌合金中氮含量
Determination of Nitrogen in Uranium and Uranium-Niobium Alloy by Optical Fiber Spectrophotometer
武红英 1李英秋 1卞敏1
作者信息
- 1. 中国工程物理研究院,四川绵阳621900
- 折叠
摘要
Abstract
By optical fiber photospectrometer, trace nitrogen in uranium and uranium-niobium alloy was determined with thymol as a chromogenic reagent. In the presence of sodium hypochlorite solution whose effective chlorine content was 0. L%-0. 1A% , ammonia was reacted with thymol to form a blue indophenol compound with maximum adsorption wavelength at 660 nm. The determination limit of this method is 20 fig/g. The relative standard deviation is less than 15%. The recovery is 88%-106/ij. These results are consistent well with that obtained with a traditional spectrophotometer.关键词
铀及铀铌合金/光纤光度计/氮Key words
uranium and uranium-niobium alloy/ optical fiber photometer/ nitrogen分类
化学化工引用本文复制引用
武红英,李英秋,卞敏..光纤光度计测定铀及铀铌合金中氮含量[J].核化学与放射化学,2012,34(2):124-128,5.