摘要

Abstract

Objective Proper process controls should be applied to decrease percentage of red blood cell spill over, to avoid loss of platelet products, and to maintain a high quality of platelet apheresis collection. Methods Using a statistical method to analyze 33 cases of red cell spill over tested out of the total 18019 platelet apheresis products in Taiyuan Blood Center from 2005 to 2009. Results Red cell spill over is related to the process of platelet apheresis collection and the maintenance condition of relevant machine. Conclusion Good pre-screening based on practical experience as a first pass, excellent technical operation to ensure whole collection process and careful maintenance of apheresis machine could help minimize the chance of red cell spill over and guarantee the effectiveness of platelets product.

关键词

机采血小板/血细胞分离机/献血者筛选

Key words

platelet apheresis/ blood cell separator/ donor screening