摘要

Abstract

A neutron ambient dose equivalent dosemeter was calibrated in 241Am-Be radionuclide neutron field.Background scattering neutrons were subtracted by adopting the shadow-cone method,semi-empirical method,polynomial fitting method and reduced-fitting method respectively.Merit and limitations of the four methods are indicated after comparison.The results are coincident with each other within the uncertainties.

关键词

中子周围剂量当量(率)仪/散射中子/校准/剂量当量响应

Key words

neutron ambient dose equivalent dosemeter/background scattering neutrons/calibration/dose equivalent response