| 注册
首页|期刊导航|桂林电子科技大学学报|局部皮肤绝热法无损测量血液灌注率研究

局部皮肤绝热法无损测量血液灌注率研究

李学亮 朱健铭 夏金虹 陈真诚

桂林电子科技大学学报2012,Vol.32Issue(4):331-334,4.
桂林电子科技大学学报2012,Vol.32Issue(4):331-334,4.

局部皮肤绝热法无损测量血液灌注率研究

Research on non-invasive measurement of blood perfusion rate with local skin adiabatic method

李学亮 1朱健铭 1夏金虹 1陈真诚1

作者信息

  • 1. 桂林电子科技大学 电子工程与自动化学院,广西桂林541004
  • 折叠

摘要

Abstract

Blood perfusion rate(BPR) of body tissue was a important parameter. However, all the currently-used measuring methods had their limitations. A non-invasive measurement of BPR was made by local skin adiabatic method based on thermal physics > and a BPR measurement system was designed. The temperature information of the skin surface collected by measuring probe was sent to PC through NI data acquisition card for data analyzing and processing, after that the value of BPR was got. The evaluation of the system was performed by clinical experiments. After further data analysis, it was found that the test data and the data from laser doppler flowmeter(LDF) is highly related, a correlation coefficient R is up to 0. 912, the proposed local skin adiabatic method in feasible for measuring BPR.

关键词

血液灌注率/局部皮肤绝热法/无损测量/NI数据采集卡

Key words

blood perfusion rate/ local skin adiabatic method/ non-invasive measurement/ NI data acquisition card

分类

信息技术与安全科学

引用本文复制引用

李学亮,朱健铭,夏金虹,陈真诚..局部皮肤绝热法无损测量血液灌注率研究[J].桂林电子科技大学学报,2012,32(4):331-334,4.

基金项目

国家863计划重点项目(2007AA02006) (2007AA02006)

桂林电子科技大学学报

1673-808X

访问量0
|
下载量0