现代检验医学杂志2012,Vol.27Issue(3):123-124,127-128,4.DOI:10.3969/j.issn.1671-7414.2012.03.042
Sysmex XE-2100血细胞分析仪白细胞分类计数的性能评价
Performance Evaluation of Leukocyte Differential Countwith Sysmex XE-2100 Automated Hematology Analyzer
马艳 1高云龙 1吴新忠 1郑松柏 1胡珺 1庞鑫1
作者信息
摘要
Abstract
Objective To evaluate the performance of leukocyte differential count with Sysmex XE-2100 automated hematology analyzer. Methods According to H20-A standard approved by National Committee for Clinical Laboratory Standards (NCCLS/CLSI),the performance of leucocyte differential count was evaluated from imprecision, comparability and clinical determination of sensitivity. Results The within-run, short-term and long-term imprecision of the instrument were between 0. 91 %~15. 53%. Compared with the manual microscopy classification, the correlation coefficient( r2 ) of neutrophil, lymphocyte, monocyte,eosinophil and basophil were 0. 956 4,0. 941 8,0. 700 7,0. 931 3 and 0. 811 1 respectively. The clinical determination of sensitivity to finding of abnormal samples were between 50%~100% in 200 clinical cases. Conclusion Sysmex XE-2100 automated hematology analyzer had good clinical determination of sensitivity,low imprecision and fine correlation with manual microscopy classification. It is suitable for the initial screening of clinical diseases.关键词
Sysmex XE-2100/白细胞分类计数/评价Key words
sysmex XE-2100/leukocyte differential count/ evaluation分类
医药卫生引用本文复制引用
马艳,高云龙,吴新忠,郑松柏,胡珺,庞鑫..Sysmex XE-2100血细胞分析仪白细胞分类计数的性能评价[J].现代检验医学杂志,2012,27(3):123-124,127-128,4.