| 注册
首页|期刊导航|物理学报|软X射线条纹相机透射式Au与CsI阴极谱响应灵敏度标定

软X射线条纹相机透射式Au与CsI阴极谱响应灵敏度标定

曾鹏 袁铮 邓博 袁永腾 李志超 刘慎业 赵屹东 洪才浩 郑雷 崔明启

物理学报2012,Vol.61Issue(15):379-385,7.
物理学报2012,Vol.61Issue(15):379-385,7.

软X射线条纹相机透射式Au与CsI阴极谱响应灵敏度标定

Spectral response calibration of Au and Csl transmission photocathodes of X-ray streak camera in a 60-5500 eV photon energy region

曾鹏 1袁铮 2邓博 2袁永腾 2李志超 2刘慎业 2赵屹东 3洪才浩 3郑雷 3崔明启3

作者信息

  • 1. 清华大学工程物理系,北京100084/中国工程物理研究院激光聚变研究中心,绵阳621900
  • 2. 中国工程物理研究院激光聚变研究中心,绵阳621900
  • 3. 中国科学院高能物理研究所,北京100049
  • 折叠

摘要

Abstract

A method is described of measuring absolute spectral response for Au and CsI transmission photocathodes in soft X-ray streak camera,which is of great importance for the inertial confinement fusion(ICF) diagnostics.Transmission photocathode is conventionally employed as photo-to-electron conversion accessories.To derive quantity information of X-ray spectra,the absolute response of photocathode must be calibrated in a range of interest.Here Au and CsI transmission photocathodes with slits are calibrated respectively on Beijing Synchrotron Radiation Facility(BSRF),in a photon energy range of 60 eV—5500 eV.This method has an uncertainty less than 10%and good feasibility.Calibration results are in good agreement with the calculation results obtained from the Henke’s photon emission model,with CH substrate effect revised.

关键词

X射线条纹相机/光阴极/同步辐射

Key words

X-ray streak camera/photocathode/BSRF

分类

数理科学

引用本文复制引用

曾鹏,袁铮,邓博,袁永腾,李志超,刘慎业,赵屹东,洪才浩,郑雷,崔明启..软X射线条纹相机透射式Au与CsI阴极谱响应灵敏度标定[J].物理学报,2012,61(15):379-385,7.

物理学报

OA北大核心CSCDCSTPCDSCI

1000-3290

访问量4
|
下载量0
段落导航相关论文