宁波市饮用水源地浮标式水质监测系统试运行问题及解决对策
Problems of Float Type Water Quality Monitoring System during Trial Running in the Sources of Drinking Water in Ningbo and Solutions
周国韧1
作者信息
- 1. 宁波市环境监测中心,浙江宁波515012
- 折叠
摘要
Abstract
During the test run of the float type water quality monitoring system located in the sources of drinking water in Ningbo,the monitoring values of pH, dissolved oxygen, conductivity, blue-green algae are accurate, while ammonia ni- trogen, nitrate nitrogen, chlorophyll are relatively poor compared with the monitoring data according the GB method. The accuracy and comparability of the seven monitoring values of the system have b.een greatly improved after enhancing the maintenance and measures of quality control. The float type water quality monitoring system can play a warning effect of water pollution of the source of drinking water in Ningbo and thus provide the protection of people's safety.关键词
浮标/水质监测系统/水库/饮用水Key words
Float type/Water quality monitoring system/Reservoir/Drinking water分类