调强放疗中多叶准直器漏射线对晶体受照剂量的影响

张玉海 李月敏 夏火生 王杰 刘亚军

中国医疗设备2012,Vol.27Issue(6):11-13,3.
中国医疗设备2012,Vol.27Issue(6):11-13,3.DOI:10.3969/j.issn.1674-1633.2012.06.003

调强放疗中多叶准直器漏射线对晶体受照剂量的影响

Impact of MLC Dose Leakage on Lens in Intensity Modulated Radiotherapy

张玉海 1李月敏 1夏火生 1王杰 1刘亚军2

作者信息

  • 1. 中国人民解放军第309医院放疗科,北京100091
  • 2. 中国人民解放军第309医院医学工程科,北京100091
  • 折叠

摘要

Abstract

Objective To study the impact of MLC dose leakage on lens in 1MRT. Methods MLC transmission factor was measured with PTW 30013 ion chamber, and the dose verification for lens of 5 IMRT plans were performed in the Topslane cylinder phantom using PTW 31010 ion chamber. Results MLC transmission factor was 1.42% and the measured doses of lens were more than calculated values (P<0.05). Conclusion MLC leakage and scattered ray on lens result in the higher dose than calculated value. It is suggested that the verification measurement of organs at risk should be performed in some cases.

关键词

调强放射疗法/多叶准直器/晶体/剂量验证

Key words

intensity modulated radiotherapy/ multi-leaf collimator/ lens/ dose verification