Objective; To investigate the optimal phase and slice thickness of double source CT venography in the head in order to get better images of cerebral veins. Methods: The small-dose tests were performed in 11 healthy volunteers and the theoretical threshold was obtained by the time-density curve. Forty-five volunteers were divided randomly into three groups to explore the virtual threshold of the bolus-triggered technique. The data obtained were retrospectively reconstructed with slice thickness of 0. 75 mm and 1. 50 mm. Results: The optimal threshold of CTV in the head was 100 HU. The quality of images with slice thickness of 0. 75 mm was superior to the quality with slice thickness of 1. 50 mm. Conclusion: Satisfactory CTV images in the head can be got with optimal threshold of 100 HU and reconstructed with slice thickness of 0. 75 mm.关键词
体层摄影术,X线计算机/静脉造影术/脑Key words
tomography, X-ray computed /phlebography/ brain分类
医药卫生引用本文复制引用
姜保东,闫华,李笃民,马祥兴,王青,张晓明,王茜,于富华..双源CT脑静脉造影扫描技术探讨[J].中国中西医结合影像学杂志,2012,10(4):305-307,3.基金项目
山东省博士基金资助项目(编号:BS2009YY036). (编号:BS2009YY036)