医疗卫生装备2012,Vol.33Issue(10):104-105,2.
HA-8160全自动糖化血红蛋白分析仪检测系统的性能验证
Performance Verification of HA-8160 Automatic Glycosylated Hemoglobin Analyzer
董磊 1刘娟 2马红雨 1全首祯 1吕建晓 1朱美财1
作者信息
- 1. 空军总医院临床检验中心,北京100142
- 2. 空军总医院输血科,北京100142
- 折叠
摘要
Abstract
Objective To verify the main analyzing performance of HA-8160 automatic glycosylated hemoglobin analyzer, which is produced by ARKRAY in Japan. Methods The precision, carryover rates and linearity of the selected testing items were evaluated. Results All the precision and carryover rates of the selected testing items were within an acceptable range. The regression equation between the theoretical and measured values in the linear verification experiment was y=bx+a, in which b value was within the range of 0.97~1.03, and a value was within an acceptable range. Conclusion The precision, carryover rates and linearity of the selected testing items of HA-8160 automatic glycosylated hemoglobin analyzer accord with the demands of
the experiments in clinical diagnosis.
关键词
检测系统/性能验证/糖化血红蛋白分析仪Key words
detection system/ performance verification/ glycosylated hemoglobin analyzer分类
医药卫生引用本文复制引用
董磊,刘娟,马红雨,全首祯,吕建晓,朱美财..HA-8160全自动糖化血红蛋白分析仪检测系统的性能验证[J].医疗卫生装备,2012,33(10):104-105,2.