摘要
Abstract
The phenomenon of size-of-source effect (SSE) is discussed and the coefficient of SSE is used for estimating the SSE property of the infrared thermometer. The experiment apparatus used for measuring SSE under lower temperatures condition is described, and several types of infrared thermometers are applied to the measurements of SSE. The experiment results indicate that the influence of SSE may be not considered at the diameters of the diaphragm hole being more than 40mm. This conclusion is suitable to the most industrial infrared thermometers.关键词
计量学/光源尺寸效应/红外温度计/校准/黑体辐射源/SSE系数Key words
Metrology/ Size-of-source effect/ Infrared thermometer/ Calibration/ Blackbody radiation source/ Coefficient of SSE分类
通用工业技术引用本文复制引用
赵泉钦,李文军,孙坚,余时帆,章龙飞,崔超..较低温度条件下SSE影响红外温度计校准的实验研究[J].计量学报,2012,33(6):509-512,4.