实用临床医药杂志2012,Vol.16Issue(22):121-123,3.
血细胞检测结果的比对和临床可接受性的评价
Comparison of blood cell analysis and clinically acceptable evaluation
周立军1
作者信息
- 1. 江苏省姜堰市人民医院,江苏姜堰,225500
- 折叠
摘要
Abstract
Objective To evaluate the comparison of blood cell cytometry results and clinical acceptability. Methods By using Sysmex XE-2100 and Beckman Coulter 5-Diff blood cell analyzers, EDTA-K2 anticoagulant whole blood samples from 40 patients and normal controls with different ranges were tested in accordance with standard operating procedures, and the results were compared. Results All the samples were checked twice and results were drawn scatter plot and bias chart. The good linear relationship between the two blood cell analyzer instruments could be visually found with small deviation and outliers. Both results were met the established quality targets. Conclusion The measurement outcomes of WBC, RBC, HGB, HCT and PLT by two blood cell analyzers in our laboratory were comparable, and the deviations were within the clinically acceptable range.关键词
血细胞检测/结果比对/可接受性评价Key words
blood cell analysis/result comparison/acceptable evaluation分类
医药卫生引用本文复制引用
周立军..血细胞检测结果的比对和临床可接受性的评价[J].实用临床医药杂志,2012,16(22):121-123,3.