Varian 23EX加速器低MU对X射线剂量学影响的研究

曾自力 林锋 王勇兵

医疗卫生装备2012,Vol.33Issue(11):117-118,121,3.
医疗卫生装备2012,Vol.33Issue(11):117-118,121,3.

Varian 23EX加速器低MU对X射线剂量学影响的研究

Dosimetry Influence of X-Ray Generated from Varian 23EX Accelerator with Low Monitor Unit

曾自力 1林锋 1王勇兵1

作者信息

  • 1. 柳州市柳铁中心医院肿瘤科,广西柳州545007
  • 折叠

摘要

Abstract

Objective To discuss the dosimetry influence of X-ray generated from Varian 23EX accelerator with low monitor unit. Methods The output dose rate was set as 300 MU/min with X-ray of Varian 23EX accelerator. X-ray of 6 and 15 MV were calibrated, then absorbed doses of low monitor unit(l~10 MU) were measured and compared with corresponding ones. The influence on start-up characteristics of accelerator with 6 and 15 MV was analyzed. Radiation quality(D20/D10) of low monitor unit was measured and compared with D20/D10 by conventional measure; percent depth dose(PDD) of low monitor unit was measured and compared with that of conventional measure. Results The influence on absorbed dose of X-ray was great at low monitor unit, with the maximum error of -11.00%; the influence on D20/D10 PDD was little, with the maximum error of 0.340% and 0.483%. Conchusion The influence on absorbed dose of the accelerator with low monitor unit from different factories and with different models and output dose rates has to be examined periodically.

关键词

加速器/MU/X射线/剂量学

Key words

accelerator, monitor unit/ X-ray/ dosimetry