郑州大学学报(医学版)2012,Vol.47Issue(6):833-836,4.DOI:10.3969/j.issn.1671-6825.2012.06.026
新双源CT Flash扫描冠状动脉CT血管成像对冠状动脉狭窄的评估效果
Assessment of dual-source CT Flash scans for coronary artery stenosis diseases
夏宾 1杨学华 1高剑波 1谭红娜 1王天佑 1梁晓雪1
作者信息
- 1. 郑州大学第一附属医院放射科,郑州,450052
- 折叠
摘要
Abstract
Aim:To evaluate the value of dual-source Flash CT angiography( Flash CTA ) in the assessment of coronary artery stenosis diseases. Methods:The results of Flash CTA and coronary artery angiography( CAG ) from 52 patients were retrospectively reviewed. Results:The diagnosis sensitivity, specificity, positive predictive value, negative predictive value and Youde index of coronary artery stenosis with Flash CTA were 84.3% ,95.7% ,89.7% ,93.2% ,and 0. 80 ,respectively. And the effective radiation dose was ( 0. 98 ~ 1. 15 ) mSv. The result of Flash CTA was well agreed with CAG( Kappa = 0. 888,P <0.001 ). Conclusion:Flash CTA with a low effective radiation dose,is valuable for the diagnosis of coronary artery stenosis,especially for moderate and severe stenosis.关键词
双源CT/冠状动脉CT血管成像/冠状动脉造影Key words
dual source CT/coronary artery CT angiography/ coronary artery angiography分类
医药卫生引用本文复制引用
夏宾,杨学华,高剑波,谭红娜,王天佑,梁晓雪..新双源CT Flash扫描冠状动脉CT血管成像对冠状动脉狭窄的评估效果[J].郑州大学学报(医学版),2012,47(6):833-836,4.