临床输血与检验2013,Vol.15Issue(1):22-24,3.DOI:10.3969/j.issn.1671-2587.2013.01.007
Sysmex XT-2000i血液分析仪血小板聚集报警信息评价
Performance Evaluation of Sysmex XT-2000i Hqematology Analyzer in Platelet Aggregation Function
林杰 1陈力平 1肖立 1何清 1程国伟1
作者信息
- 1. 214044无锡,中国人民解放军第101医院检验科
- 折叠
摘要
Abstract
Objective To evaluate sysmex XT-2000i automated haematology analyzer in analysis of platelet aggregation according to the "PLT lumps?" and measurement of maximum platelet aggregation rate. Methods 345 patients were analyzed by sysmex XT-2000i and plasma maximum platelet aggregation rate (MAR) was detected. The results were compared and analyzed statistically. Results Platelet aggregation rate of positive group judged by sysmex XT-2000i was significant higher than negative group(p<0. 01), and the two methods of evaluating platelet aggregation function had certain dependability (Kappa = 0. 61). Conclusion The information proved in the platelet detecting process of sysmex XT-2000i may provide certain basis to preliminary analysis judgment of the platelet aggregation function.关键词
血细胞分析/血小板聚集Key words
Haematology analyze/Platelet aggregation分类
医药卫生引用本文复制引用
林杰,陈力平,肖立,何清,程国伟..Sysmex XT-2000i血液分析仪血小板聚集报警信息评价[J].临床输血与检验,2013,15(1):22-24,3.